Atellica® VTLi Patient-side Immunoassay Analyzer

Specification

When a patient enters the emergency department presenting with the symptoms of a myocardial infarction (MI), every moment is critical. Achieving favorable clinical and operational outcomes requires fast and accurate chest pain assessment at the point of care, as well as coordination with the laboratory for quality assurance. Having high-sensitivity troponin testing at the point of care—producing accurate results in just 8 minutes—could transform care delivery for millions of people each year, improving ED throughput, reducing stress and strain on acute care clinicians, and aiding in optimizing the use of time and resources of busy laboratories.

Benefits

Simple to use and easily integrated into existing systems, the Atellica® VTLi Patient-side Immunoassay Analyzer, powered by Magnotech® Technology, delivers accurate, clinically actionable high-sensitivity troponin I results directly in the hands of those who need it most.

Technical Details

Technical Specifications

Sample Type

Capillary (fingerstick) whole blood

 

Lithium-heparinized venous whole blood

 

 

Lithium-heparinized venous plasma

 

Sample Size

30–100 μL

Analysis

Time to Results: ~8 minutes

Quality Control

Sero PATHONORM Cardiac Acute Liq

Reagent Cartridge

Storage: 2–8°C

Size of box (W x D x H): 152 x 95 x 167 mm

Cartridges per box: 24

System Dimensions

Length: 25 cm

Height: 5.2 cm

Width: 8.5 cm

Weight: 780 g

Color LCD display: 7.3 inches

Integrated Barcode Scanner

Patient ID, Operator ID, and bottled QC; 1D (linear) and 2D barcode symbologies

Operating System

Microsoft WEC2013

Data Security

Patient data anonymization

TLS 1.2 encrypted end-to-end communication, using client and server certificates

Configurable level of user authentication

No storage of passwords, no hard-coded passwords

No user access to operating system

Share this